Cdi-torque MULTITEST Government User Manual

Browse online or download User Manual for Equipment Cdi-torque MULTITEST Government. CDI Torque MULTITEST Government User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 96
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
USER MANUAL
MANUAL P/N 2000-760-02
MULTITEST
Electronic Torque/Force/Angle
Calibration System
TFC2000-0-03
NSN 6635-01-327-4461
Page view 0
1 2 3 4 5 6 ... 95 96

Summary of Contents

Page 1 - MULTITEST

USER MANUALMANUAL P/N 2000-760-02MULTITESTElectronic Torque/Force/Angle Calibration SystemTFC2000-0-03NSN 6635-01-327-4461

Page 2

Using This ManualSafety MessagesSafety messages are provided to help prevent personal injuryand equipment damage. All safety messages are introduced

Page 3 - 3.75 - 50 in lb

ISafety InformationIMPORTANT SAFETY INSTRUCTIONSThis manual contains important safety and operatinginstructions for CDI MULTITEST Torque calibration

Page 4 - Table of Contents

IISafety InformationWARNINGRisk of electric shock and fire.• For indoor use only. Do not expose charger torain or snow. Do not use in damp locations

Page 5

IIISafety InformationWARNING• Make sure all components, including, adaptors,extensions, drivers and sockets are rated tomatch or exceed the torque or

Page 6

IVSafety Information

Page 7 - Table of Illustrations

1-1IntroductionThe MULTITEST Indicator is a laboratory grade instrument thatprovides TORQUE, FORCE and ANGLE of rotationmeasurements and set-able aler

Page 8

1-2Introductio nSet-up and calibration programming is entered on front panelmembrane keys. HIGH and LOW torque limits are adjustable togive an audibl

Page 9 - Using this Manual

2-1Description andSpecificationsThe MULTITEST Torque Calibration System is comprised of anIndicator, Mechanical Loader, Roll Cabinet, Transducers and

Page 10 - ! CAUTION

2-2Functional Description and SpecificationsB — Transducer (Accessory)The loader accepts all MULTITEST series single transducersdirectly. Use a 2000-

Page 11 - Safety Information

2-3Functional Description and SpecificationsMULTITEST IndicatorThis section describes the major hardware components of theMULTITEST Indicator, includi

Page 12

19220 San Jose Avenue • City of Industry, California 91748-1497 (626) 965-0668 / (800) 525-6319 • Office Fax (626) 810-2759 • Sales Fax (626) 965-241

Page 13

2-4Functional Description and SpecificationsF — Maximum Range DisplayFigure 2-4: Maximum Range DisplayTwo line by 16 character 5x8 dot-matrix LCD disp

Page 14

2-5Functional Description and SpecificationsG — Front Panel Membrane Function KeysToggles between TORQUE and FORCE measurement modes.Selects TRACK, P

Page 15 - Introduction

2-6Functional Description and SpecificationsOpens and terminates calibration mode.Edits DATE/TIME programming function.Sets up high torque/force limit

Page 16 - Introductio n

2-7Functional Description and SpecificationsRear Panel (Input/Output)The MULTITEST Indicator rear panel is shown in Figure 2-5.Figure 2-5: 2000-610-0

Page 17 - Functional

2-8Functional Description and SpecificationsD — PRINTER PORT (DB-9P)PIN FUNCTION2 Receive3 Transmit5 GroundE — OPTIONAL (DB-9P)PIN FUNCTION2 Receive3

Page 18 - D — Roll Cabinet

2-9Functional Description and SpecificationsMULTITEST Torque/Force TransducersMULTITEST torque/force transducers provide industry standardinternal squ

Page 19 - Front Panel

2-10Functional Description and SpecificationsDisplay Resolution for Transducers in Force ModeDisplay Resolution for Transducers in Torque ModeMULTITES

Page 20 - F — Maximum Range Display

2-11Functional Description and SpecificationsSpecificationsMULTITEST Torque Calibration SystemSystem Accuracy±0.25% of reading @ 25°C(Indicator and tr

Page 21

2-12Functional Description and SpecificationsData Storage/Recall w/Date Time Stamp3,000 measurementsStatistical AnalysisMax, Min, Range, Mean, Sigma N

Page 22

2-13Functional Description and SpecificatonsTorque/Force TransducersAccuracy±0.2% of reading @ 25°C, within specified range, when used asprescribed wi

Page 23 - Rear Panel (Input/Output)

Suggested Transducer Calibration Check Points, Bars, Hangers and Weightsfor 2500-400-02 (Multiple Transducer Assembly, 0.25% Accuracy) Bar Hanger Cal

Page 24 - G — VOLTAGE SELECT

2-14Functional Description and Specifications

Page 25

3-1Setup andProgrammingMULTITEST Torque Calibration System SetupFigure 3-1: MULTITEST Torque Tester/Calibrator/Loader SetupA – Internal printerB – In

Page 26

3-2Setup and ProgrammingSetting up the MULTITEST Torque Calibration System1. Position the roll cabinet on a level floor and lock thewheels.2. Mount th

Page 27

BAC DEFG3-3Setup and ProgrammingMULTITEST Indicator ControlsRefer to the illustration below when performing the power-upand programming procedures.Fig

Page 28

3-4Setup and ProgrammingPowering the EquipmentTo power the equipment, press the power button on the front ofthe MULTITEST Indicator.• At POWER ON, the

Page 29 - Torque/Force Transducers

3-5Setup and ProgrammingProgramming SetupSetting up Date and Time1. Press DATE/TIME. The LCD display readsDate dd/month/yy - upper lineTime hh:mm:ss

Page 30

3-6Setup and ProgrammingSetting Up High and Low Limits AlertTorque/Force ModeUse SET HIGH and SET LOW to protect the tool under test orserve as a fast

Page 31 - Programming

3-7Setup and ProgrammingSetting High Torque/Force Limit1. Press SET HIGH.— The LED display reads:last SET HIGH value , with one digit flashing— The LC

Page 32 - Setup and Programming

3-8Setup and ProgrammingAUTO CLEAR can be programmed with a 1 to 9 second delay.Use this feature to visually note the reading before the displayclears

Page 33 - MULTITEST Indicator Controls

3-9Setup and ProgrammingDATA LIST MemoryWhen changing sensors, or when creating a new statistical datalist, the data list memory must be cleared. Th

Page 34 - 4-in-1 Transducer Select

Table of ContentsMULTITEST Indicator Manual ...

Page 35 - Programming Setup

3-10Setup and Programming4. Press ENTER to clear the last memory location. Thenumber of total memory locations decrements by one.Pushing ENTER repeat

Page 36

3-11Setup and ProgrammingClear All Memory LocationsUse the procedure in this section to clear all memory locations.1. Press RECALL DATA.— LCD displays

Page 37 - Programming AUTO CLEAR

3-12Setup and Programming3. Press ENTER.— The LED display reads:PRINTER or PC— The LCD display reads:Download: - upper linePort - lower line4. Press u

Page 38

3-13Setup and Programming8. Press right DIGIT.— LED display indicates:YES or NO— LCD display reads:Download: - upper lineSend Init yes/no - lower line

Page 39 - DATA LIST Memory

3-14Setup and ProgrammingB. Rotate the encoder shaft exactly one complete turn(360 degrees). The upper right LCD display readsthe total number of cou

Page 40 - Clear Single Memory Location

Setup and Programming3-15Setting Up Torque-Angle Measurement ModesIn using the torque-angle or "turn-of-the-nut" algorithm, afastener is run

Page 41 - Selecting Printer/PC Ports

3-16Setup and Programming3. Using up and down VALUE and left and right DIGIT,enter the clamping torque from the fastener torque-anglespecification ont

Page 42

Setup and Programming3-175. Press SET HIGH.— The LED display shows the previously programmeddegrees of rotation preset value.— The LCD display reads:

Page 43 - Encoders

3-18Setup and Programming

Page 44

4-1Using the MultitestTorque Calibration SystemThe 2000-610-02 Indicator is designed to work specifically withthe MULTITEST series of smart transducer

Page 45 - Set Up Torque-Angle Algorithm

Table of ContentsSafety Information ...

Page 46

4-2Using the MULTITEST Toque Calbration SystemDial and Bending Beam Torque Wrenches and ScrewdriversBe sure the MULTITEST transducer is capable of han

Page 47

4-3Using the MULTITEST Toque Calbration SystemExample:When tool accuracy is 4% and MULTITEST Indicator /Transducer System accuracy is 0.25%, readings

Page 48

4-4Using the MULTITEST Toque Calbration System5. Remove the tool from the transducer-loader and waitapproximately 30 seconds then press ZERO TARE. 6.

Page 49 - Torque Calibration System

4-5Using the MULTITEST Torque Calbration SystemTesting Power ToolsThe dynamic torque characteristic of a power tool and the statictorque applied when

Page 50 - Screwdrivers

4-6Using the MULTITEST Toque Calbration System3. Press MODE to select TRACK mode. The green TRACKlamp lights.4. Adjust the power tool to the desired

Page 51

4-7Using the MULTITEST Toque Calbration System13. While holding the power tool drive perfectly in-line withthe transducer drive, reinstall the tool on

Page 52

4-8Using the MULTITEST Toque Calbration SystemUse the following procedure to make force measurements.1. To program the MULTITEST Indicator, refer to

Page 53 - Non-Impact Tools

4-9Using the MULTITEST Toque Calbration SystemDisplaying Statistical Analysis on LCD1. Press STAT.— The LED display reads:** STAT**— The LCD displays

Page 54

4-10Using the MULTITEST Toque Calbration SystemPrinting Statistical Analysis1. Press STAT.— The LED display reads:**STAT**— The LCD display reads:HIT

Page 55

4-11Using the MULTITEST Toque Calbration SystemDisplaying or Downloading Data ListTorque measurements stored in the data list also include a dateand t

Page 56

iiTable of ContentsDownloading Data List to Printer ... 4-11Downloadin

Page 57 - Figure 4-3: Value Display

4-12Using the MULTITEST Toque Calbration System5. Press ENTER to download the raw data list. The LEDdisplay reads:SENDING6. Press ENTER again to retu

Page 58 - Printing Statistical Analysis

4-13Using the MULTITEST Toque Calbration SystemHow to Use Analog Output (Rear Panel BNCConnector)Analog Output on the MULTITEST Indicator provides a r

Page 59 - Displaying Data List on LCD

4-14Using the MULTITEST Toque Calbration SystemMeasuring Torque and AngleAlthough measuring torque and angle is not a torque testingfeature, transduce

Page 60

4-15Using the MULTITEST Toque Calbration SystemAngle OnlyThe MULTITEST accepts input from common bi-directionalincremental rotary encoders. The Indic

Page 61 - Connector)

4-16Using the MULTITEST Toque Calbration SystemTorque-Angle AlgorithmIn addition to MULTITEST series torque transducers, theMULTITEST Indicator accept

Page 62 - Torque Only

5-1CalibrationThe MULTITEST Torque Calibration system is factory calibratedusing precision torque test arms, certified weights and laboratorygrade equ

Page 63 - Angle Only

5-2Calib rationMounting DetailsSingle Transducer Mounting Detail for CalibrationSingle transducer calibration requires the use of a reactionstand and

Page 64 - Torque-Angle Algorithm

5-3CalibrationTorque Over 250 ft lbFor torque calibration above 250 ft lb, the calibration block ismoved to the left for CCW loading and to the right

Page 65 - Calibration

5-4Calib ration2. Remove separated adaptor from calibration block.3. Attach transducer to the calibration block so that redmark on transducer lines up

Page 66 - Mounting Details

5-5CalibrationQuick CheckAlways be alert to the potential for personal injurythat may be caused by excessive torqueapplications, careless handling of

Page 67

Table of IllustrationsFunctional Description and Specifications ... 2-1Figure

Page 68 - Calib ration

5-6Calib rationSuggested Transducer Calibration Check Points, Bars, Hangers and Weights(Sheet 1 of 2)Part Bar Hanger Calibration Check Weights to Add

Page 69 - Quick Check

5-7CalibrationSuggested Transducer Calibration Check Points, Bars, Hangers and Weights(Sheet 2 of 2)Part Bar Hanger Calibration Check Weights to Add T

Page 70

5-8Calib rationTorque/Force CalibrationsMULTITEST Torque/Force Transducer Calibration Improper calibration can result in torquemeasurement errors. Fo

Page 71

5-9Calibration5. Enter the following code into the display using theDIGIT right and left keys and the up and down VALUEkeys:CODE52226. Press ENTER.— T

Page 72 - Torque/Force Calibrations

5-10Calib ration14. Apply and release full scale torque to the transducerthree times in the CW direction, then remove the weightsand hanger. Leave th

Page 73

5-11Calib rationMULTITEST Indicator Torque/Force Calibration Improper calibration can result in torquemeasurement errors. Follow these proceduresprec

Page 74

5-12Calib ration7. Program the measured voltage (example provided in step5 on the previous page) into the MULTITEST Indicatorusing the up and down VAL

Page 75 - IMPORTANT

5-13CalibrationGravitational EffectsCorrection Factors on Test WeightsWeights used in torque calibration are affected by theiracceleration due to:• Al

Page 76

5-14Calib rationFree Air Correction for AltitudeCorrection Factor for Check Point Readings1. Determine latitude and altitude. Obtain information from

Page 77 - Gravitational Effects

6-12000-600-02Manual LoaderThe 2000-600-02 Manual Loader is designed as an integralcomponent of the CDI MULTITEST Torque Calibration Systemproviding a

Page 79 - Manual Loader

6-22000-600-02 Motorized LoaderTorque/force transducers, purchased separately, are available inranges from 15-200 in oz, to 200-2000 ft lb. The MULTI

Page 80 - Functional Description

6-32000-600-02 Manual LoaderTorque Tester/CalibratorThe 2000-600-02 Manual Loader mounts on a CDI roll cabinet,part number 2000-100-02. An indicator

Page 81 - 2000-600-02 Manual Loader

6-42000-600-02 Manual LoaderUse the 2000-600-02 Manual Loader to test torque wrenches.Refer to Figure 6-2 for an exploded view of the Loader andcompo

Page 82 - B — Hand Crank

6-52000-600-02 Manual LoaderD — Linear BearingsMay be slipped over the reaction pins to reduce friction whentesting very small or high resolution elec

Page 83

6-62000-600-02 Manual LoaderLoader Rear Panel ControlsThe 2 0 00-6 0 0 -02 M a nual Loader r ear p a nel i s sho w n in Figure6-3. Figure 6-3

Page 84 - Loader Rear Panel Controls

6-72000-600-02 Manual LoaderBall Handle AdaptorWrenches featuring a ball handle are fixtured using the ballhandle reaction adaptor shown in Figure 6-

Page 85 - Ball Handle Adaptor

6-82000-600-02 Manual Loader4-in-1 Transducer (Accessory)The 4-in-1 transducer is mounted to the loader using thestandoff fixture and bracket asse

Page 86 - 4-in-1 Transducer (Accessory)

6-92000-600-02 Manual LoaderExtension Arm (Accessory)The 2000-600-02 Manual Loader accepts torque wrenches up to80" in length. The extension arm

Page 87 - Extension Arm (Accessory)

6-102000-600-02 Manual LoaderTypes of Loader TestingTorque Driver TestingAn accessory torque screwdriver testing kit, 2000-750-02, isavailable for use

Page 88 - Types of Loader Testing

6-112000-600-02 Manual LoaderForce TestingThe 2000-600-02 Manual Loader and MULTITEST seriestransducers can be configured to deliver compression andte

Page 89 - Force Testing

vUsing this ManualThis manual contains instructions for use and setup of theMULTITEST Torque Calibration System. A table of contents and atable of

Page 90

6-122000-600-02 Manual Loader

Page 91 - Accessories

A-1AppendixAccessoriesThis appendix contains information on replacement parts andaccessories that are available from Consolidated Devices,Inc.,1-800-5

Page 92

A-2AccessoriesMULTITEST Torque Calibration SystemComponents and Inventory ControlComponent Description Drawer-Cutout2000-100-02 Roll cabinet2000-100-2

Page 93

A-3AccessoriesOptional Transducers and AccessoriesComponent Description Drawer-Cutout2000-5-021/4" ext. drive15-200 in oz transducer 9-B1/4"

Page 94

A-4AccessoriesComponent Description Drawer-Cutout2000-550-02 Extension arm kit, for up to 80" wrenches2000-550-108 Arm2000-550-115 Reaction pin

Page 95 - Calibration Accessories

A-5AccessoriesCalibration AccessoriesCalibration of the 2000-610-02 Indicator and MULTITEST seriestransducers is recommended at least every 1 year.Com

Page 96

A-6AccessoriesComponent Description Drawer-CutoutWeights2500-310-1 0.3 oz Hook (1) 62500-310-2 0.5 oz Hook (1) 62000-310-2 2 oz Bob (1) 62000-310-3 4

Comments to this Manuals

No comments