Setup and Programming
Clear All Memory Locations
Use the procedure in this section to clear all memory locations.
1. Press RECALL DATA.
LCD displays as shown in Figure 3-5. The present
memory location at the upper left, the total number
of memory locations to the right and the time data
was stored at far right.
LED display shows the measurement value of that
memory location while the LCD displays its
engineering units at the lower right.
2. Press MANUAL CLEAR.
LCD display shows "CLEAR" and the present (left)
and total (right) memory locations on the upper line,
as shown in Figure 3-6.
3. Press right DIGIT twice to select "All" on the lower LCD
line.
4. Press ENTER to clear the entire data memory. LED
display reads:
NO DATA
The audible alarm sounds momentarily. The 2000-610-
02 automatically returns to measurements mode.
Setting Up Printer/PC Ports
The MULTITEST Indicator is equipped with two RS-232 serial
communications ports for use with an external printer or
personal computer. The transfer rate is programmable from 300
to 19.2K baud rate with 8 data bits, 1 stop bit and no parity.
Select "PRINTER" to enable the printer port.
The built-in printer is connected to the PRINTER port.
To disable the built-in printer, use the
INTERNAL/EXTERNAL switch on the back of the 2000-
610-02 Indicator.
Select "PC" to enable the OPTIONAL COM port.
Selecting Printer/PC Ports
1. Press CAL to enter calibration mode.
LED display reads the following with the least
significant digit (LSD) flashing:
CODE0000
2. Use left and right DIGIT to select the appropriate digits
and up and down VALUE to increment or decrement
each digit between "0" and hexidecimal "F," program the
number "2000" into the LED display.