5-9
Calibration
5. Enter the following code into the display using the
DIGIT right and left keys and the up and down VALUE
keys:
CODE5222
6. Press ENTER.
The LED display reads:
SELECT
The LCD display reads:
Calibrate CW
7. Press ENTER.
The LED display reads a flashing number in the
LSD.
The LCD display read:
Enter no. of transducers
8. Using the up and down VALUE keys, enter 1for a
single transducer or4 for the 4-in-1 transducer.
If you enter "1" skip to step 11.
9. Press ENTER. The LED display reads a flashing number
in the LSD and the LCD display reads:
Select transducer
10. Using the up and down VALUE keys, enter the number of
the 4-in-1 transducer to calibrate. Possible values are:
1 = 50 in oz
2 = 400 in lb
3 = 1000 in lb
4 = 250 ft lb
11. Press ENTER.
The LED display banners:
ENTER FS
followed by the last full scale calibration point for the
particular transducer installed or selected with its
MSD flashing.
The LCD display reads:
ENTER FS transducer
[units]
12. Using the up and down VALUE keys and the left and
right DIGIT keys, enter the full scale torque capacity of
the transducer in the engineering units selected.
13. Press ENTER.
The LED display banners:
ZERO TARE
Four zeros follow shifting alternately across the
display.
The LCD display reads:
CALIBRATION
<ZEROTARE> CONT.