3-9
Setup and Programming
DATA LIST Memory
When changing sensors, or when creating a new statistical data
list, the data list memory must be cleared. The entire memory
list or any single memory location may be selectively cleared.
Clearing from the top of the list decrements the total number of
memory locations. Clearing from inside the list shifts all
subsequently taken readings down from that location and
decrements the total number of memory locations.
Clear Last Memory Location
Use the procedure in this section to clear the last memory
location.
1. Press RECALL DATA. The LCD displays as shown in
Figure 3-5. The present memory location is at the upper
left, the total number of memory locations to its right and
the time the data was stored at the far right.
The LED display shows the measurement value of
that memory location.
The LCD displays its engineering units at the lower
right.
Figure 3-5: RECALL Data Display
2. Press MANUAL CLEAR. LCD display shows "CLEAR"
and the present (left) and total (right) memory locations
on the upper line, as shown in Figure 3-6.
Figure 3-6: RECALL Data Display
3. Press right DIGIT to select "Single" on the lower LCD
line.